vol.35 número2¿Puede el índice de masa corporal pregestacional relacionarse con el estado psicológico y físico de la madre durante todo el embarazo?El método BFMNU como alternativa a los métodos en uso basados en energía: estudio de la correlación entre la energía alimentaria y la masa corporal índice de autoresíndice de assuntospesquisa de artigos
Home Pagelista alfabética de periódicos  

Serviços Personalizados

Journal

Artigo

Indicadores

Links relacionados

  • Em processo de indexaçãoCitado por Google
  • Não possue artigos similaresSimilares em SciELO
  • Em processo de indexaçãoSimilares em Google

Compartilhar


Nutrición Hospitalaria

versão On-line ISSN 1699-5198versão impressa ISSN 0212-1611

Resumo

COSTA-MOREIRA, Osvaldo; PATROCINIO-DE-OLIVEIRA, Cláudia-Eliza  e  DE-PAZ, José-Antonio. Dual energy X-ray absorptiometry (DXA) reliability and intraobserver reproducibility for segmental body composition measuring. Nutr. Hosp. [online]. 2018, vol.35, n.2, pp.340-345. ISSN 1699-5198.  https://dx.doi.org/10.20960/nh.1295.

Background:

the measurement precision of body composition can interfere on the diagnosis and prescription of diseases' treatment. Furthermore, with regard to dual energy X-rays absorptiometry (DXA), there may be important differences between the measures of regions of interest (ROI) automatically performed by DXA or manually by an evaluator, which can cause measurement error and influence the evaluation or diagnosis.

Aim:

thus, this study aimed to evaluate the measurement reliability of body composition by DXA and intraobserver reproducibility for the ROI measurement.

Methods: a total of 15 young adults were subjected to two full-body scans by DXA, under the same conditions. The first scan of all volunteers was chosen and the extent of the ROI was performed, in two stages, separated by a period of 2 weeks, by a single trained evaluator. The coefficient of variation (CV) and intraclass correlation coefficient (ICC) were calculated with a significance level of p < 0.05.

Results:

high-precision measurements of DXA for whole body mass (ICC = 0.999; CV = 2.3%), fat mass (ICC = 0.998; CV = 1.6%), lean mass (ICC = 0.995; CV = 0.3%) and bone mineral content (ICC = 0.996; CV < 0.1%) were obtained. Further, it was observed high intraobserver reproducibility for ROI measurement, with ICC values ranging between 0.952 and 0.999.

Conclusion:

body composition measurement by DXA presents high reproducibility for whole body mass, fat mass, lean mass and bone mineral content and also high intraobserver reproducibility for the ROI measurement.

Palavras-chave : Densitometry; Reliability of results; Reproducibility of results; Body composition; Bone mineral density.