vol.36 número4Variabilidad en los criterios de definición y métodos de detección de la disfunción renal en las unidades de cuidados intensivos: ¿se aplican los consensos internacionales para el diagnóstico de la disfunción renal?Resumen del documento de consenso «Guías de práctica clínica para el manejo del síndrome de bajo gasto cardiaco en el postoperatorio de cirugía cardiaca» índice de autoresíndice de assuntospesquisa de artigos
Home Pagelista alfabética de periódicos  

Serviços Personalizados

Journal

Artigo

Indicadores

Links relacionados

  • Em processo de indexaçãoCitado por Google
  • Não possue artigos similaresSimilares em SciELO
  • Em processo de indexaçãoSimilares em Google

Compartilhar


Medicina Intensiva

versão impressa ISSN 0210-5691

Resumo

CANABAL, A. et al. Foreseeable variation in parameters measured at implant and follow-up of permanent pacemaker active fixation electrodes. Med. Intensiva [online]. 2012, vol.36, n.4, pp.270-276. ISSN 0210-5691.

Objectives: To analyze the variations in the parameters relative to active fixation electrodes at the time of implantation and over subsequent follow-up during 6 months of the acute phase of implantation. Design: A descriptive, analytical, prospective, observational cohort study was made of consecutive cases over a period of 8 months (April-December 2010). Setting: Pacing unit of an Intensive Care Unit. Patients or participants: Patients undergoing permanent pacemaker implantation with active fixation electrodes, implanted in both atrium and ventricle, Interventions: Measurement of variables described with a threshold analyzer during electrode fixation and at different times during the study, Main compared variables: threshold, impedance and intrinsic activity (both atrial and ventricular) before and after fixation, at 48hours, at one month and 6 months, Comparisons were made using the Student t-test for paired data, assuming significance for p<0,05, and ANOVA to analyze the successive changes over ambulatory follow-up. Results: We analyzed 40 patients, with 19 atrial and 40 ventricular electrodes, In fixation, the electrodes showed significant variation in the impedance values of the atrial lead (1,188,53±397,26 vs 610,69±326,30 ohms, p<0,0001) and ventricular lead (1,512,93±718,07 vs 768,80±224,90 ohms, p>0,0001), In the first 48hours it was coupled with a decrease in ventricular (0,86±0,35 vs 0,48±0,23 volts, p=0,0001) and atrial pacing threshold (1,10±0,39 vs 0,43±0,23 volts, p=0,0003), and p-wave sensing (3,61±2,25 vs 2,32±1,09mV, p=0,0463), Over follow-up we found the parameters to be stable, with no significant changes. Conclusions: After active lead fixation, a fall in impedance of the antrial and ventricular is expected, Over the next 48hours improvement in atrial and ventricular threshold may occur, in contrast to the sensitivity of the intrinsic activity, which reached significance at the P wave measured after 48hours, These values stabilize over patient follow-up and do not differ significantly in the studied acute patient course.

Palavras-chave : Lead impedance; Pacemaker; Active fixation.