SciELO - Scientific Electronic Library Online

 
vol.21 issue1Scholrstooth decay ratio from Barbastro Primary Care Service author indexsubject indexarticles search
Home Pagealphabetic serial listing  

Services on Demand

Journal

Article

Indicators

Related links

  • On index processCited by Google
  • Have no similar articlesSimilars in SciELO
  • On index processSimilars in Google

Share


Avances en Odontoestomatología

On-line version ISSN 2340-3152Print version ISSN 0213-1285

Abstract

JODAR, S et al. Use of ionizing radiations in spanish dental equipments: intraoral and panoramic. Av Odontoestomatol [online]. 2005, vol.21, n.1, pp.361-370. ISSN 2340-3152.

The aim of the study was to determinate Dental Radiology behaviour in Murcia Community to control the European Union Recommendations. 1822 official reports on quality assurance in intraoral radiology and 155 official reports on panoramic radiology have been studied. The results indicate that one and a half of intraoral radiographic equipments studied and 6´24% in panoramic had alterations. Only the 63´54% of the surgeries achived the European Community Recommendations of intraoral radiology working to 70 kVp, 8 mA and 1.5 mm of Al. There have been a decrease of the intraoral radiation dose exposured. The recent legislation have meant a decrease in the intraoral radiation dose administrated despite of the unachivement of European Community Recommendations.

Keywords : Radiodiagnosis; dental radiography; mean doses; assurance quality.

        · abstract in Spanish     · text in Spanish     · Spanish ( pdf )

 

Creative Commons License All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License