Mi SciELO
Servicios Personalizados
Revista
Articulo
Indicadores
- Citado por SciELO
- Accesos
Links relacionados
- Citado por Google
- Similares en SciELO
- Similares en Google
Compartir
Actas Urológicas Españolas
versión impresa ISSN 0210-4806
Resumen
ALONSO RODRIGUEZ, Diego et al. Follow-up of the FlowSecure artificial urinary sphincter. Actas Urol Esp [online]. 2009, vol.33, n.9, pp.952-955. ISSN 0210-4806.
The AMS-800 artificial urinary sphincter has been the only prosthesis available for treatment of stress urinary incontinence refractory to other therapeutic modalities for the past 25 years. The relatively high rate of complications occurring with the AMS-800 device during this time led to introduce a number of cahnges in its design that resulted in a new prostheses, the FlowSecure artificial sphincter. The FlowSecure artificial urinary sphincter is an adjustable prosthesis filled with normal saline without contrast. Plain X-rays cannot therefore be used for monitoring, and ultrasound is the most adequate radiographic technique for evaluation. In addition to calculating the post-void residue, ultrasound allows for verifying prosthesis status and for calculating the urethral occluding pressure. A detailed clinical history and flow rate measurement should be used together with the ultrasound scan to functionally assess patients with the FlowSecure device in order to determine the need for adjusting system pressure to the minimum pressure required for total continence.
Palabras clave : Follow-up; FlowSecure; Ultrasound scan.