SciELO - Scientific Electronic Library Online

 
vol.33 número5Cirugía compleja en el tratamiento con implantes del maxilar superiorCarga precoz con sobredentadura mandibular sobre dos implantes insertados mediante cirugía guiada índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

Links relacionados

  • En proceso de indezaciónCitado por Google
  • No hay articulos similaresSimilares en SciELO
  • En proceso de indezaciónSimilares en Google

Compartir


Avances en Odontoestomatología

versión On-line ISSN 2340-3152versión impresa ISSN 0213-1285

Resumen

MATOS GARRIDO, N et al. Influence of the cyclic fatigue in the implant-abutment internal connections. Av Odontoestomatol [online]. 2017, vol.33, n.5, pp.221-229. ISSN 2340-3152.

Introduction:

The connection implant-abutment has been suggested to be crucial for the long-term success of osseointegration and to prevent future complications. The aim of this study was to evaluate the influence of fatigue cyclic test in the behavior of internal connection implant-abutment.

Methods:

Fifty titanium implants with internal connections were divided in 3 groups: octogonal connection (OC); double hexagonal connection (DHC); and hexagonal connection (HC) and with their respective abutments. Abutments were tightened to 30 Ncm with a torque controller. A cyclic load between 550 N and 135N according to different implant systems at a 30-degree angle to the long axis was applied to the implants for a 5 million cycles.

Results:

Biomechanical testing showed implant fracture in a limit load of 325 N of OC implants; in 150 N of DHC implants and 135 N of HC implants. Fracture cracks were located in the area between neck and body of implants. No abutments and screws retention fractures were found.

Conclusions:

Results of this study indicate that fatigue cyclic tests are important to evaluate the biomechanical behavior of different connection implants-abutments of several implants systems.

Palabras clave : Dental implants; cyclic loading; fatigue test; biomechanics; implant-abutment connection; implant dentistry.

        · resumen en Español     · texto en Español     · Español ( pdf )