vol.80 número2Influencia de la aberración esférica en la función visual tras cirugía de catarata: ensayo prospectivo aleatorioPrevalencia de la retinopatía diabética en la población de diabéticos diagnosticados en las comarcas de Girona: Estudio de los factores asociados índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

Links relacionados

Compartir


Archivos de la Sociedad Española de Oftalmología

versión impresa ISSN 0365-6691

Resumen

MORENO MONTANES, J; SADABA, LM  y  HERAS, H. Intraocular pressure measurement using the Proview® tonometer. Arch Soc Esp Oftalmol [online]. 2005, vol.80, n.2, pp.79-83. ISSN 0365-6691.

Purpose: To evaluate a new ocular tonometer (Proview®) which functions by visualizing a phosphene after putting pressure on the upper eyelid. To ascertain its accuracy and reproducibility with respect to the Goldmann tonometer (GT). Methods: A study on both eyes of 110 non-selected patients was performed. One measurement with GT and three subsequent measurements with Proview® were taken by the same investigator. The number of failed attempts to visualize the phosphene was recorded. We evaluated each eye separately to observe the possible learning effect. Results: The intraocular pressure (IOP) mean with Proview® is 5 mm Hg higher than the GT (p<0.001). The coefficient of correlation between both instruments is low (R = 0.67; R2=0.45) in the right eye (RE). The difference between maximum and minimum values of the 3 measurements taken with the Proview® was 3.21 (SD: 1.73) mm Hg in RE. The failed attempts to see the phosphene were 5.14 (SD: 2.39) in RE and 4.93 (SD: 2.78) in left eye (p>0.05). Conclusions: The Proview® tonometer showed low accuracy and reproducibility in comparison with the GT. This tonometer requires a long learning process before phosphene visualization. The results demonstrated that this tonometer is not clinically useful, except in patients with serious corneal diseases which make measurement with GT very difficult.

Palabras clave : Intraocular pressure; phosphene tonometry; glaucoma.