<?xml version="1.0" encoding="ISO-8859-1"?><article xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance">
<front>
<journal-meta>
<journal-id>0210-5691</journal-id>
<journal-title><![CDATA[Medicina Intensiva]]></journal-title>
<abbrev-journal-title><![CDATA[Med. Intensiva]]></abbrev-journal-title>
<issn>0210-5691</issn>
<publisher>
<publisher-name><![CDATA[Elsevier España, S.L.]]></publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id>S0210-56912009000600003</article-id>
<title-group>
<article-title xml:lang="es"><![CDATA[Trabajo y patrón respiratorio con diferentes niveles de ventilación asistida proporcional]]></article-title>
<article-title xml:lang="en"><![CDATA[Respiratory work and pattern with different proportional assist ventilation levels]]></article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author">
<name>
<surname><![CDATA[Ruiz-Ferrón]]></surname>
<given-names><![CDATA[Francisco]]></given-names>
</name>
<xref ref-type="aff" rid="A01"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname><![CDATA[Machado]]></surname>
<given-names><![CDATA[Juan]]></given-names>
</name>
<xref ref-type="aff" rid="A01"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname><![CDATA[Morante]]></surname>
<given-names><![CDATA[Antonia]]></given-names>
</name>
<xref ref-type="aff" rid="A01"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname><![CDATA[Galindo]]></surname>
<given-names><![CDATA[Silvia]]></given-names>
</name>
<xref ref-type="aff" rid="A01"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname><![CDATA[Castillo]]></surname>
<given-names><![CDATA[Ana]]></given-names>
</name>
<xref ref-type="aff" rid="A01"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname><![CDATA[Rucabado]]></surname>
<given-names><![CDATA[Luis]]></given-names>
</name>
<xref ref-type="aff" rid="A01"/>
</contrib>
</contrib-group>
<aff id="A01">
<institution><![CDATA[,Complejo Hospitalario de Jaén Hospital Médico-Quirúrgico Unidad de Medicina Intensiva]]></institution>
<addr-line><![CDATA[Jaén ]]></addr-line>
<country>España</country>
</aff>
<pub-date pub-type="pub">
<day>00</day>
<month>09</month>
<year>2009</year>
</pub-date>
<pub-date pub-type="epub">
<day>00</day>
<month>09</month>
<year>2009</year>
</pub-date>
<volume>33</volume>
<numero>6</numero>
<fpage>267</fpage>
<lpage>268</lpage>
<copyright-statement/>
<copyright-year/>
<self-uri xlink:href="http://scielo.isciii.es/scielo.php?script=sci_arttext&amp;pid=S0210-56912009000600003&amp;lng=en&amp;nrm=iso"></self-uri><self-uri xlink:href="http://scielo.isciii.es/scielo.php?script=sci_abstract&amp;pid=S0210-56912009000600003&amp;lng=en&amp;nrm=iso"></self-uri><self-uri xlink:href="http://scielo.isciii.es/scielo.php?script=sci_pdf&amp;pid=S0210-56912009000600003&amp;lng=en&amp;nrm=iso"></self-uri><abstract abstract-type="short" xml:lang="es"><p><![CDATA[Objetivo: Estudiar el grado mínimo de asistencia, en ventilación proporcional asistida (VPA), que disminuya el trabajo respiratorio a límites fisiológicos (0,6 j/l), y la relación entre los cambios en el patrón respiratorio y el esfuerzo que realiza el paciente a diferentes niveles de VPA. Diseño: estudio de cohortes prospectivo. Ámbito: Unidad de Medicina Intensiva polivalente del Complejo Hospitalario de Jaén. Pacientes y métodos: Se incluye a 12 pacientes que han precisado ventilación mecánica prolongada, y cumplen criterios para iniciar la retirada del respirador. Intervenciones: utilizamos el respirador Puritan-Bennett 840 en modo proporcional asistido. De forma aleatoria se modifica el porcentaje de asistencia entre el 5 y el 80%, en intervalos del 10%. Antes de cada modificación del nivel de VPA, el paciente recibe ventilación asistida controlada por volumen seguida de presión soporte. Principales variables de interés: medimos la mecánica respiratoria antes del inicio de VPA y, en este modo de ventilación asistida, el patrón respiratorio y el trabajo respiratorio. Resultados: El descenso en la asistencia respiratoria con VPA conlleva un aumento significativo del trabajo respiratorio, que va desde 0,2 ± 0,07 (0,1-0,3) j/l con VPA80 hasta 0,9 ± 0,2 (0,4-1,5) j/l con VPA5 (p = 0,002). El coeficiente de correlación entre el nivel de VPA y el trabajo respiratorio (medido en j/l y j/min) fue r = -0,8 y -0,6, respectivamente. El grado mínimo de asistencia en relación con un trabajo fisiológico fue del 30% (0,63 ± 0,13 j/l). Excepto el volumen corriente que aumentó de forma significativa (VPA80 de 0,4 ± 0,1; VPA5 de 0,3 ± 0,1; p = 0,02), los demás parámetros que definen el patrón respiratorio no cambiaron con el aumento de VPA. Conclusiones: En el grupo de pacientes estudiado, al aumentar la asistencia respiratoria con VPA disminuye el trabajo respiratorio, sin cambiar el patrón respiratorio de forma significativa. Valores menores del 30% de VPA conllevan un excesivo trabajo respiratorio.]]></p></abstract>
<abstract abstract-type="short" xml:lang="en"><p><![CDATA[Objective: To study the minimum assistance level during proportional assist ventilation (PAV) to decrease the work of breathing to physiological limits (0.6 j/l) and the relationship between breathing pattern changes and respiratory effort at different PAV levels. Design: Prospective cohort study. Setting: Polyvalent intensive care unit of a teaching hospital of Jaen, Spain. Patients and methods: Twelve long-term mechanical ventilated patients who met criteria to initiate weaning from the ventilator. Interventions: We used the Puritan-Bennett 840 ventilator in proportional assist ventilation. The percentage of support was randomly modified between 5% and 80%, in intervals of 10%. Prior to the change in the PAV level, the patients were ventilated in assist-volume control followed by pressure support ventilation. Main variables of interest: Before PAV, we measured the respiratory mechanics and the breathing pattern and work of breathing during this mode. Results: The decrease in respiratory assist in PAV was related to significantly higher work of breathing, this going from 0.2 ± 0,07 (0.1-0.3) j/l with PAV80 to 0.9 ± 0.2 (0.4-1.5) j/l with PAV5 (p = 0.002). The coefficient correlation between PAV level and work of breathing (measured as j/l and j/min) was r = -0.8 and -0.6, respectively. Minimum PAV level related with physiological work of breathing was 30% (0.63 ± 0.13 j/l). Except for the tidal volume that increased significantly (PAV80 vs PAV5 = 0.4 ± 0.1 vs 0.3 ± 0.1; p = 0.02), the remaining variables defining the breathing pattern did not changed with the increase in PAV. Conclusions: In the group of patients studied, the increase in the PAV levels decreases work of breathing, without significantly changing the breathing pattern. Levels lower than 30% of PAV are associated to excessive work of breathing.]]></p></abstract>
<kwd-group>
<kwd lng="es"><![CDATA[Ventilación mecánica]]></kwd>
<kwd lng="es"><![CDATA[Interacción paciente-respirador]]></kwd>
<kwd lng="es"><![CDATA[Ventilación proporcional asistida]]></kwd>
<kwd lng="es"><![CDATA[Trabajo respiratorio]]></kwd>
<kwd lng="en"><![CDATA[Mechanical ventilation]]></kwd>
<kwd lng="en"><![CDATA[Patient-ventilator interactions]]></kwd>
<kwd lng="en"><![CDATA[Proportional assist ventilation]]></kwd>
<kwd lng="en"><![CDATA[Work of breathing]]></kwd>
</kwd-group>
</article-meta>
</front><body><![CDATA[ <p><a name="top"></a><font face="Verdana" size="2"><b>ORIGINAL</b></font></p>     <p>&nbsp;</p>     <p><font face="Verdana" size="4"><b>Trabajo y patrón respiratorio con diferentes niveles de ventilación asistida proporcional</b></font></p>     <p><font face="Verdana" size="4"><b>Respiratory work and pattern with different proportional assist ventilation levels</b></font></p>     <p>&nbsp;</p>     <p>&nbsp;</p>     <p><font face="Verdana" size="2"><b>Francisco Ruiz-Ferrón, Juan Machado, Antonia Morante, Silvia Galindo, Ana Castillo y Luis Rucabado</b></font></p>     <p><font face="Verdana" size="2">Servicio de Cuidados Críticos y Urgencias, Unidad de Medicina Intensiva, Hospital Médico-Quirúrgico, Complejo Hospitalario de Jaén, Jaén, España</font></p>     <p><font face="Verdana" size="2"><a href="#bajo">Dirección para correspondencia</a></font></p>     <p>&nbsp;</p>     ]]></body>
<body><![CDATA[<p>&nbsp;</p> <hr size="1">    <p><font face="Verdana" size="2"><b>RESUMEN</b></font></p>     <p><font face="Verdana" size="2"><b>Objetivo:</b> Estudiar el grado mínimo de asistencia, en ventilación proporcional asistida (VPA), que disminuya el trabajo respiratorio a límites fisiológicos (0,6 j/l), y la relación entre los cambios en el patrón respiratorio y el esfuerzo que realiza el paciente a diferentes niveles de VPA. Diseño: estudio de cohortes prospectivo. Ámbito: Unidad de Medicina Intensiva polivalente del Complejo Hospitalario de Jaén.    <br><b>Pacientes y métodos:</b> Se incluye a 12 pacientes que han precisado ventilación mecánica prolongada, y cumplen criterios para iniciar la retirada del respirador. Intervenciones: utilizamos el respirador Puritan-Bennett 840 en modo proporcional asistido. De forma aleatoria se modifica el porcentaje de asistencia entre el 5 y el 80%, en intervalos del 10%. Antes de cada modificación del nivel de VPA, el paciente recibe ventilación asistida controlada por volumen seguida de presión soporte. Principales variables de interés: medimos la mecánica respiratoria antes del inicio de VPA y, en este modo de ventilación asistida, el patrón respiratorio y el trabajo respiratorio.    <br><b>Resultados:</b> El descenso en la asistencia respiratoria con VPA conlleva un aumento significativo del trabajo respiratorio, que va desde 0,2 ± 0,07 (0,1-0,3) j/l con VPA80 hasta 0,9 ± 0,2 (0,4-1,5) j/l con VPA5 (p = 0,002). El coeficiente de correlación entre el nivel de VPA y el trabajo respiratorio (medido en j/l y j/min) fue r = -0,8 y -0,6, respectivamente. El grado mínimo de asistencia en relación con un trabajo fisiológico fue del 30% (0,63 ± 0,13 j/l). Excepto el volumen corriente que aumentó de forma significativa (VPA80 de 0,4 ± 0,1; VPA5 de 0,3 ± 0,1; p = 0,02), los demás parámetros que definen el patrón respiratorio no cambiaron con el aumento de VPA.    <br><b>Conclusiones:</b> En el grupo de pacientes estudiado, al aumentar la asistencia respiratoria con VPA disminuye el trabajo respiratorio, sin cambiar el patrón respiratorio de forma significativa. Valores menores del 30% de VPA conllevan un excesivo trabajo respiratorio.</font></p>     <p><font face="Verdana" size="2"><b>Palabras clave:</b> Ventilación mecánica; Interacción paciente-respirador; Ventilación proporcional asistida; Trabajo respiratorio.</font></p> <hr size="1">    <p><font face="Verdana" size="2"><b>ABSTRACT</b></font></p>     <p><font face="Verdana" size="2"><b>Objective:</b> To study the minimum assistance level during proportional assist ventilation (PAV) to decrease the work of breathing to physiological limits (0.6 j/l) and the relationship between breathing pattern changes and respiratory effort at different PAV levels. Design: Prospective cohort study. Setting: Polyvalent intensive care unit of a teaching hospital of Jaen, Spain.    <br><b>Patients and methods:</b> Twelve long-term mechanical ventilated patients who met criteria to initiate weaning from the ventilator. Interventions: We used the Puritan-Bennett 840 ventilator in proportional assist ventilation. The percentage of support was randomly modified between 5% and 80%, in intervals of 10%. Prior to the change in the PAV level, the patients were ventilated in assist-volume control followed by pressure support ventilation. Main variables of interest: Before PAV, we measured the respiratory mechanics and the breathing pattern and work of breathing during this mode.    ]]></body>
<body><![CDATA[<br><b>Results:</b> The decrease in respiratory assist in PAV was related to significantly higher work of breathing, this going from 0.2 ± 0,07 (0.1-0.3) j/l with PAV80 to 0.9 ± 0.2 (0.4-1.5) j/l with PAV5 (p = 0.002). The coefficient correlation between PAV level and work of breathing (measured as j/l and j/min) was r = -0.8 and -0.6, respectively. Minimum PAV level related with physiological work of breathing was 30% (0.63 ± 0.13 j/l). Except for the tidal volume that increased significantly (PAV80 vs PAV5 = 0.4 ± 0.1 vs 0.3 ± 0.1; p = 0.02), the remaining variables defining the breathing pattern did not changed with the increase in PAV.    <br><b>Conclusions:</b> In the group of patients studied, the increase in the PAV levels decreases work of breathing, without significantly changing the breathing pattern. Levels lower than 30% of PAV are associated to excessive work of breathing.</font></p>     <p><font face="Verdana" size="2"><b>Key words:</b> Mechanical ventilation; Patient-ventilator interactions; Proportional assist ventilation; Work of breathing.</font></p> <hr size="1">     <p>&nbsp;</p>     <p><font face="Verdana" size="2"><b>Introducción</b></font></p>     <p><font face="Verdana" size="2">La ventilación asistida es necesaria en pacientes con impulso respiratorio normal o aumentado, que no pueden mantener una adecuada respiración espontánea por debilidad muscular y/o anormal mecánica respiratoria, su empleo representa hasta el 50% del tiempo total de soporte ventilatorio<sup>1</sup>. Es fundamental en ventilación asistida una adecuada interacción entre el paciente y el respirador, punto con frecuencia subestimado<sup>2</sup>, y que puede prolongar la retirada definitiva de este último<sup>3</sup>. La ventilación proporcional asistida (VPA) es un modo de ventilación diseñado para mejorar la sincronía con el respirador<sup>4</sup>, aunque tiene otras utilidades, como el estudio de la estabilidad del centro respiratorio<sup>5</sup>. Este modo de ventilación no impone un patrón respiratorio, sino que se programa un porcentaje de asistencia, por lo tanto, es el propio centro respiratorio del paciente lo que determina el patrón de ventilación más adecuado<sup>6</sup>, con lo que mejora la interacción con el respirador<sup>7</sup>. Para ello, se precisa medir la mecánica respiratoria del paciente, y estimar las demandas de flujo y volumen<sup>8</sup>.</font></p>     <p><font face="Verdana" size="2">Durante la VPA es necesario programar un grado de asistencia que evite esfuerzo o reposo excesivos por el riesgo tanto de fatiga como de atrofia muscular, respectivamente<sup>2</sup>. Aunque habitualmente las variables que definen el patrón respiratorio se utilizan para valorar el esfuerzo que realiza el paciente, esta aproximación en pacientes críticos es, al menos, cuestionable<sup>9</sup>. El objetivo de nuestro estudio es determinar, en VPA, el mínimo grado de asistencia respiratoria relacionado con un trabajo respiratorio mayor que el fisiológico, y los cambios en el patrón respiratorio con diferentes grados de asistencia respiratoria.</font></p>     <p>&nbsp;</p>     <p><font face="Verdana" size="2"><b>Pacientes y métodos</b></font></p>     <p><font face="Verdana" size="2">Durante un periodo de 3 meses, se estudió a un grupo de pacientes que precisaban ventilación mecánica, ingresados en una unidad de medicina intensiva polivalente de un hospital de segundo nivel. El protocolo de estudio fue aprobado por el comité de investigación del hospital e informamos al paciente y/o familiar de su realización.</font></p>     ]]></body>
<body><![CDATA[<p>&nbsp;</p>     <p><font face="Verdana" size="2"><b>Pacientes</b></font></p>     <p><font face="Verdana" size="2">Se incluyó a 12 pacientes que precisaban ventilación mecánica por fallo respiratorio agudo de diferentes etiologías (<a href="#t1">tabla 1</a>), en situación clínica estable y con criterios de inicio de soporte respiratorio parcial, pero no para realizar una prueba de desconexión del respirador. La sedación se había retirado al menos 24 horas antes, pero se mantuvo aún un valor residual de sedación de 3 puntos en la escala de Ramsay. Posición en decúbito supino con el cabecero elevado 30<sup>o</sup>. En todos los pacientes la intubación fue endotraqueal translaríngea con un diámetro de tubo endotraqueal entre 7,5 y 8,5 mm. Se monitorizó de forma continua la frecuencia cardíaca, la frecuencia respiratoria, la presión arterial y la pulsioximetría, sin utilizar ninguna instrumentación adicional.</font></p>     <p align=center><font face="Verdana" size="2"><a name="t1"><img src="/img/revistas/medinte/v33n6/original1_t1.jpg" width="600" height="247"></a></font></p>     <p>&nbsp;</p>     <p><font face="Verdana" size="2"><b>Protocolo de ventilación</b></font></p>     <p><font face="Verdana" size="2">Todos los pacientes recibieron ventilación mecánica con un respirador Puritan-Bennett 840 (Tyco, Gosport, Reino Unido), equipado con el modo VPA. Éste realiza un análisis automático<sup>10,11</sup> de la elastancia (EstPAV+) y las resistencias espiratorias (RsrPAV+). Al inicio de la VPA, estas mediciones se realizan en 4 ciclos consecutivos. El primer nivel de asistencia se realiza considerando la resistencia de la vía aérea artificial y un valor de resistencia y distensibilidad (Csr) ajustado al peso del paciente. Luego, el respirador, de forma automática, repite las mediciones al azar a intervalos de 4-10 ciclos respiratorios. Mediante oclusiones de 300 ms al final de la inspiración, mide la presión meseta (P<sub>meseta</sub>) para el cálculo de la distensibilidad considerando el valor de PEEP (Csr = volumen corriente &#091;Vc&#093; / P<sub>meseta</sub> - PEEP). Esta oclusión no interfiere con el patrón respiratorio espontáneo, ya que el aumento en TI mecánico evoca una débil respuesta neuromuscular<sup>12</sup>. Las resistencias espiratorias totales se miden en la curva de flujo espiratorio tras una oclusión, utilizando el valor medio de 3 puntos (máximo flujo y tras 5 y 10 ms). El respirador considera las resistencias in vitro del tubo endotraqueal utilizado (Rtet), lo que permite el cálculo de las resistencias del sistema respiratorio (Rsr<sub>PAV</sub> = Rtot - Rtet). En el modo VPA se programa el porcentaje de descarga (k, máximo del 95%) y el respirador genera una presión según la ecuación de movimiento, considerando la resistencia no lineal del tubo endotraqueal, Paw<sub>(t)</sub> = K &#091;V'I<sub>(t)</sub>× (Rtet<sub>(t)</sub> + Rrs<sub>PAV</sub>) + V<sub>(t)</sub> × Ers<sub>PAV</sub>&#093;. Donde Paw<sub>(t)</sub> es la presión en la vía aérea en un momento determinado, Rtet es la resistencia del tubo endotraqueal dependiente del flujo. V'I<sub>(t)</sub> es el flujo inspiratorio y V<sub>(t)</sub> es el volumen corriente.</font></p>     <p><font face="Verdana" size="2">El respirador muestra el trabajo respiratorio calculado, basado en la monitorización no invasiva de la presión muscular<sup>13</sup>. Este método asume que la presión total aplicada al sistema respiratorio (P<sub>total</sub>) es la suma de la presión generada por el respirador (Pres) y la realizada por la musculatura respiratoria del paciente (Pmus), así P<sub>total</sub> = Pres + Pmus. Si la presión derivada de la inercia es mínima, la presión total es la suma de la disipada en las resistencias (Rsr) y las propiedades elásticas del sistema respiratorio (Esr), éstas son función del flujo (V') y el volumen (V) sobre capacidad residual funcional. De este modo, P<sub>total</sub> = V' × Rsr + V × Esr. Midiendo la presión total aplicada al sistema, la elastancia y las resistencias, así como el flujo y el volumen (medidos cada 5 ms), se puede calcular la presión muscular. La integración de la presión muscular en función del volumen permite calcular el trabajo respiratorio realizado.</font></p>     <p><font face="Verdana" size="2">Inicialmente, la ventilación asistida se controló por volumen con flujo constante (AC); la sensibilidad del <i>trigger</i>, por flujo al mínimo nivel sin relación con autociclado, y la PEEP se ajustó al 80% del nivel de auto-PEEP estática medida. Calculamos la mecánica respiratoria (distensibilidad, resistencias y auto-PEEP estática) durante la ventilación mecánica controlada con flujo constante, según la técnica de oclusión con los dispositivos del respirador. Seguidamente, iniciamos la ventilación en presión de soporte, con un grado inicial de asistencia igual a la presión meseta del modo AC, sin modificar la FiO<sub>2</sub> ni el valor de PEEP. El nivel de soporte de presión se disminuyó progresivamente hasta que el paciente presentara un impulso respiratorio, sin signos de excesivo trabajo respiratorio (sudoración, utilización de musculatura accesoria, taquicardia, taquipnea, etc.). En ese momento el paciente pasa a VPA; tras un periodo de estabilización, anotamos el trabajo y el patrón respiratorios (volumen corriente, frecuencia respiratoria, volumen minuto, flujo inspiratorio medio, relación tiempo inspiratoriotiempo total del ciclo respiratorio), utilizando el valor medio de 20 ciclos respiratorios. Se modificó, de forma aleatoria, entre el 5 y el 80% el nivel de VPA, en pasos del 10%. Para comenzar la ventilación proporcional asistida en similar estado basal, antes de cada nivel de VPA el paciente pasa a ventilación AC, modificando el patrón de ventilación hasta quedar dependiente del respirador. Previamente al nuevo nivel de VPA se reinicia la ventilación en presión soporte, y se disminuye el grado de asistencia progresivamente. Por lo tanto, el tiempo de cada fase de ventilación y el protocolo completo es variable de un paciente a otro. Después, todos los pacientes quedaron en ventilación con el nivel mínimo de VPA relacionado con un trabajo respiratorio dentro del intervalo clínicamente aceptable (<U>&lt;</U> 0,6 j/l)<sup>14</sup>.</font></p>     <p><font face="Verdana" size="2">Analizamos los resultados mediante pruebas no paramétricas y regresión lineal, utilizamos el test de Friedman para comparación de medias. Los datos se expresan como media ± desviación estándar y media (intervalo de confianza de la media del 95%). Consideramos como estadísticamente significativo un valor de p &lt; 0,05.</font></p>     ]]></body>
<body><![CDATA[<p>&nbsp;</p>     <p><font face="Verdana" size="2"><b>Resultados</b></font></p>     <p><font face="Verdana" size="2">Las características de la mecánica respiratoria y el patrón de ventilación se muestran en la <a href="#t2">tabla 2</a>.</font></p>     <p align=center><font face="Verdana" size="2"><a name="t2"><img src="/img/revistas/medinte/v33n6/original1_t2.jpg" width="600" height="282"></a></font></p>     <p>&nbsp;</p>     <p><font face="Verdana" size="2">El trabajo respiratorio calculado disminuye de forma significativa con el aumento en la asistencia respiratoria (<a href="#f1">fig. 1</a>). Así, con VPA5 es de 0,9 ± 0,2 j/l y con VPA80, 0,2 ± 0,07 j/l (p = 0,002). Medido durante 1 min el trabajo respiratorio también es significativamente menor (VPA5-80, 8,6 ± 4,4 frente a 2,05 ± 0,9 j/min; p = 0,002). El coeficiente de correlación entre el grado de asistencia y el trabajo respiratorio (expresado en j/l frente a j/min) es -0,8 (r = -0,6).</font></p>     <p align=center><font face="Verdana" size="2"><a name="f1"><img src="/img/revistas/medinte/v33n6/original1_f1.jpg" width="600" height="500"></a></font></p>     <p align=center><font face="Verdana" size="2"><b>Figura 1 </b>Gráfico de caja de la relación entre el trabajo respiratorio y el nivel    <br> de ventilación proporcional asistida (VPA+).</font></p>     <p>&nbsp;</p>     ]]></body>
<body><![CDATA[<p><font face="Verdana" size="2">El nivel óptimo de VPA relacionado con un trabajo dentro del límite fisiológico (0,63 ± 0,13 j/l) fue del 30% de compensación del trabajo respiratorio, pero hay diferencias entre pacientes y, en algunos casos, niveles bajos de asistencia respiratoria disminuyen el trabajo respiratorio de forma significativa. Así, por ejemplo, en 3 pacientes con niveles de asistencia tan bajos como el 10%, el trabajo respiratorio está en intervalo fisiológico (<a href="#f1">fig. 1</a>).</font></p>     <p><font face="Verdana" size="2">Analizamos las variables que definen el patrón respiratorio a diferentes niveles de VPA. Si bien entre el grado mínimo de asistencia y el máximo (VPA5-80) los cambios en el volumen corriente son significativos (0,3 ± 0,1 l frente a 0,4 ± 0,1 l; p = 0,02), el resto de los parámetros no cambian significativamente (figs.  <a href="#f2">2</a> y <a href="#f3">3</a>). Frecuencia respiratoria, 28 ± 10 frente a 26 ± 10 rpm; volumen minuto, 9,8 ± 2,1 frente a 11,2 ± 5,1 l/min; Vt/TI, 0,47 ± 0,14 frente a 0,52 ± 0,16 l/s; TI/T<sub>total</sub>, 0,37 ± 0,06 frente a 0,35 ± 0,07.</font></p>     <p align=center><font face="Verdana" size="2"><a name="f2"><img src="/img/revistas/medinte/v33n6/original1_f2.jpg" width="600" height="543"></a>    <br><b>Figura 2</b> Cambios en el patrón respiratorio con diferentes valores de ventilación proporcional asistida    <br>(VPA+). FR: frecuencia respiratoria; TI: tiempo inspiratorio; Vc: volumen corriente.    <br>Resultados expresados como media (intervalo de confianza del 95%).</font></p>     <p align=center>&nbsp;</p>     <p align=center><font face="Verdana" size="2"><a name="f3"><img src="/img/revistas/medinte/v33n6/original1_f3.jpg" width="600" height="270"></a>    <br><b>Figura 3</b> Registro representativo del paciente 12, en el que monitorizamos la presión esofágica. Curvas    <br>de presión en vía aérea, esofágica, flujo y volumen durante la ventilación proporcional asistida (VPA+)    ]]></body>
<body><![CDATA[<br>al 80 y el 5%. Destacamos entre ambos registros el diferente esfuerzo (presión esofágica) con cambios    <br>no significativos en el patrón respiratorio. Entre ambos grados de asistencia (el 80 frente al 5%), la frecuencia    <br>respiratoria es de 30 y 36 rpm y el volumen corriente de 450 y 300 ml. En el registro inicial podemos observar    <br>oclusiones intermitentes al final de la inspiración para mediciones automáticas de mecánica respiratoria.</font></p>     <p>&nbsp;</p>     <p><font face="Verdana" size="2"><b>Discusión</b></font></p>     <p><font face="Verdana" size="2">Nuestros resultados indican que al inicio de la VPA, un grado de asistencia del 30% es suficiente para evitar un excesivo trabajo respiratorio y que los cambios en el patrón respiratorio son poco útiles para inferir el trabajo respiratorio.</font></p>     <p><font face="Verdana" size="2">Aunque el primer estudio clínico fue publicado en 1992<sup>15</sup>, este modo de ventilación no es de uso habitual. Posiblemente por la dificultad en medir la mecánica respiratoria en ventilación asistida, pues los valores en ventilación controlada no son útiles<sup>16</sup>. El método ideal sería utilizar un catéter esofágico<sup>17</sup>, pero pocas unidades disponen del equipo necesario y estas mediciones pueden no ser un perfecto indicador de la presión muscular respiratoria en ventilación asistida<sup>18</sup>. Se ha intentado reproducir en ventilación controlada el patrón respiratorio espontáneo para medir la mecánica respiratoria<sup>19</sup>, esto es complicado y precisa sedación profunda. Una alternativa práctica es aumentar progresivamente la asistencia hasta provocar sobrecompensación <i>(runaway)</i><sup>20</sup>, pero es necesario tener experiencia con este modo de ventilación. Por otro lado, aun con medidas precisas, en respiración asistida la mecánica respiratoria no es constante, sino que puede cambiar y producir escasa o excesiva asistencia. Por lo tanto, es aconsejable un sistema que mida con cierta frecuencia estos parámetros<sup>21</sup>. El respirador que utilizamos mide de forma automática e intermitente la mecánica respiratoria<sup>10</sup>, esto simplifica este modo de ventilación<sup>2</sup>. Las limitaciones de este sistema pueden deberse al análisis de las resistencias. El valor espiratorio puede ser mayor que el inspiratorio, especialmente en pacientes con limitación al flujo espiratorio<sup>22</sup>. Además, si bien el <i>software</i> VPA considera la no linealidad en la relación presión/flujo debida al tubo endotraqueal, las resistencias reales pueden ser diferentes de las teóricas<sup>23</sup>.</font></p>     <p><font face="Verdana" size="2">El grado de asistencia que disminuye el esfuerzo muscular respiratorio varía en diferentes estudios. Coincidimos con resultados que indican un esfuerzo excesivo con valores de asistencia menores del 31 ± 3% de la elastancia y el 55 ± 5% de las resistencias<sup>24</sup>. Ranieri et al<sup>25</sup>, con una asistencia del 40%, disminuyen el producto presión/tiempo a valores normales (10,7 ± 44,7 cmH<sub>2</sub>O × s<sup>-1</sup>)<sup>26</sup>, y Grasso et al<sup>19</sup>, con valores similares y de asistencia resistiva (44 ± 3) y elástica (58 ± 4). Delaere et al<sup>7</sup> describen que una asistencia del 50% es insuficiente para reducir el trabajo respiratorio a intervalo fisiológico. Esta variabilidad ha sido destacada en otros estudios, con valores entre el 32 y el 60%, si bien el valor mínimo parece estar alrededor del 40%<sup>20</sup>. Esta dispersión puede deberse a la fase evolutiva del fallo respiratorio, la mecánica respiratoria, la fuerza muscular y las demandas ventilatorias<sup>20</sup>. Así, en la fase aguda del fallo respiratorio, son necesarios valores de asistencia más altos (62,9 ± 12,7%)<sup>27</sup>. Nuestro estudio también muestra esta variabilidad, en algunos casos con una asistencia del 10%, el trabajo respiratorio calculado puede ser normal. Esto implica que el modo de ventilación debe ser individualizado en cada paciente; de acuerdo con estudios sobre modalidades de ventilación asistida que demuestran que es más importante cómo se utiliza el modo de ventilación que el modo de ventilación en sí<sup>28</sup>.</font></p>     <p><font face="Verdana" size="2">En la práctica, la asistencia respiratoria se programa según valoración clínica y variables como el volumen y la frecuencia respiratoria. Pero estos parámetros per se no son un buen indicador del trabajo respiratorio<sup>20</sup>. Aunque hay datos contradictorios, y un aumento progresivo de la asistencia en presión soporte clásicamente se ha relacionado con mejor ventilación y menor frecuencia respiratoria<sup>29</sup>. Por otro lado, se ha demostrado que el patrón respiratorio puede ser independiente de la carga mecánica y estar más relacionado con las propiedades mecánicas del sistema<sup>30</sup>, enfermedad subyacente y factores inherentes al paciente<sup>20</sup>. Nuestros resultados están de acuerdo con esta aproximación y con estudios en VPA, que demuestran que si bien el esfuerzo respiratorio disminuye de forma significativa, el patrón respiratorio se modifica poco en este modo de ventilación<sup>15,20</sup>. Incluso una sobrecarga mecánica en VPA no cambia significativamente el patrón respiratorio<sup>19</sup>. Esto contrasta con estudios en presión de soporte<sup>29</sup>; sin embargo, el estudio de Giannouli et al<sup>24</sup> muestra que el aumento progresivo de la asistencia en presión de soporte disminuye los ciclos asistidos, pero con un aumento de esfuerzos ineficaces, mientras que en VPA la frecuencia respiratoria no se modifica de forma significativa. Esto indica una diferente interacción paciente-respirador en cada modo de ventilación. Además, el patrón respiratorio en pacientes dependientes del respirador está influido por otros factores relacionados con la enfermedad subyacente, parámetros de ventilación<sup>31</sup> y efectos residuales de la sedación<sup>32</sup>. Por lo tanto, parece necesaria una correcta valoración del esfuerzo inspiratorio para pautar adecuadamente el grado de asistencia<sup>33</sup>.</font></p>     ]]></body>
<body><![CDATA[<p><font face="Verdana" size="2">En nuestro estudio, se ha calculado el trabajo respiratorio. Este análisis no necesita dispositivos adicionales, no es invasivo y tiene una buena correlación con el método estándar<sup>34</sup>, como se ha demostrado con diferentes valores de ventilación con presión soporte<sup>34,35</sup>. Las limitaciones de este análisis se deben a que asume que la mecánica respiratoria es lineal y no considera el trabajo dependiente de la auto-PEEP. En pacientes dependientes del respirador estos parámetros (E y R) cambian según el grado de asistencia respiratoria<sup>33</sup>. Esta limitación posiblemente disminuya si se considera la resistencia del tubo endotraqueal y con mediciones periódicas de estas variables, procedimientos que realiza este respirador. Por otro lado, no considerar el trabajo dependiente de la auto-PEEP puede subestimar el trabajo elástico<sup>36</sup>. Sin embargo, en la práctica no podemos pensar en mediciones exactas ya que, independientemente del método de medida del trabajo respiratorio, éste puede no estimar adecuadamente el gasto energético de la respiración, debido a la contracción tónica de la musculatura respiratoria, la distorsión de la pared torácica, la rarefacción del gas durante la inspiración o la compresión en espiración, factores que se ven favorecidos por una anormal mecánica respiratoria, taquipnea y una inadecuada interacción entre el paciente y el respirador<sup>37</sup>.</font></p>     <p><font face="Verdana" size="2">Con este modo de ventilación, los esfuerzos ineficaces no desaparecen por completo, están favorecidos por la asincronía espiratoria<sup>24</sup> y auto-PEEP. En VPA hay un retraso inherente a sensado, procesamiento de señales e inicio de la ventilación. Factores que dependen del tipo de respirador, la constante de tiempo del sistema respiratorio y el grado de asistencia programado<sup>38</sup>. Más importante son los altos valores de auto-PEEP, ya que el flujo inspiratorio y el volumen instantáneo son los factores determinantes de la presión generada por el respirador<sup>8</sup>. Por lo tanto, el esfuerzo que contrarresta la PEEPi no es sensado, medido ni asistido por el respirador. Además, una valoración inadecuada de la mecánica respiratoria, por limitación al flujo espiratorio, favorece el <i>runaway</i><sup>39</sup>. Sin embargo, la VPA se ha utilizado eficazmente en pacientes con EPOC agudizada<sup>39</sup>. Uno de los pacientes que estudiamos presentaba agudización de EPOC y recibió ventilación adecuada.</font></p>     <p><font face="Verdana" size="2">En resumen, los resultados de nuestro estudio muestran que una asistencia respiratoria de más del 30% es suficiente al inicio de la retirada de la VPA. Parece necesario utilizar parámetros directamente indicativos del trabajo respiratorio en lugar de las variables derivadas del patrón de ventilación.</font></p>     <p>&nbsp;</p>     <p><font face="Verdana" size="2"><b>Bibliografía</b></font></p>     <!-- ref --><p><font face="Verdana" size="2">1. Esteban A, Alía I, Ibañez J, Benito S, Tobin MJ and Spanish lung failure collaborative group. Modes of mechanical ventilation and weaning. A national survey of Spanish hospitals. Chest. 1994;106:1188-93.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874666&pid=S0210-5691200900060000300001&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">2. Racca F, Squadrone V, Ranieri MV. Patient-vetilator interaction during the triggering phase. Respir Care Clin. 2005;11:225-45.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874667&pid=S0210-5691200900060000300002&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">3. Thille AW, Rodríguez P, Cabello B, Lellouche F, Brochard L. Patient-ventilator asynchrony during assisted mechanical ventilation. Intensive Care Med. 2006;32:1515-22.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874668&pid=S0210-5691200900060000300003&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">4. Gay PC, Hess DR, Hill NS. Noninvasive proportional assist ventilation for acute respiratory insufficiency. Comparison with pressure support ventilation. Am J Respir Crit Care Med. 2001;164:1606-11.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874669&pid=S0210-5691200900060000300004&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">5. Meza S, Younes M. Ventilatory stability during sleep studied with proportional assist ventilation (PAV). Sleep. 1996;19: S166.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874670&pid=S0210-5691200900060000300005&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">6. Bigatello LM, Nishimura M, Imanaka H, Hess D, Kimball WR, Kacmarek RM. Unloading of the work of breathing by proportional assist ventilation in a lung model. Crit Care Med. 1997; 25:267-72.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874671&pid=S0210-5691200900060000300006&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">7. Delaere S, Roeseler J, D'hoore W, Matte P, Reynaert M, Jolliet P, et al. Respiratory muscle workload in intubated, spontaneous breathing without COPD: pressure support vs proportional assist ventilation. Intensive Care Med. 2003;29:949-54.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874672&pid=S0210-5691200900060000300007&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">8. Younes M. Proportional assist ventilation, a new approach to ventilatory support. Theory. Am Rev Respir Dis. 1992;145:114-20.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874673&pid=S0210-5691200900060000300008&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">9. Banner MJ, Kirby RR, Kirton OC, Dehaven B, Blanch PB. Breathing frequency and pattern are poor predictors of work of breathing in patients receiving pressure support ventilation. Chest. 1995;108:1338-44.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874674&pid=S0210-5691200900060000300009&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">10. Younes M, Kun J, Masiowski B, Webster K, Roberts D. A method for non-invasive determination of inspiratory resistance during proportional assist ventilation. Am J Respir Crit Care Med. 2001;163:829-39.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874675&pid=S0210-5691200900060000300010&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">11. Younes M, Webster K, Kun J, Roberts D, Masiowski T. A method for measuring passive elastance during proportional assist ventilation. Am J Respir Crit Care Med. 2001;164:50-60.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874676&pid=S0210-5691200900060000300011&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">12. Younes M, Kun J, Webster K, Roberts D. Response of ventilator-dependent patients to delayed opening of exhalation valve. Am J Respir Crit Care Med. 2002;166:21-30.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874677&pid=S0210-5691200900060000300012&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">13. Younes M. Proportional assist ventilation. En: Tobin MJ, editor. Principles and practice of mechanical ventilation. New York: MacGraw-Hill; 1994. p. 349-69.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874678&pid=S0210-5691200900060000300013&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">14. Otis AB. The work of breathing. En: Fenn WO, Rahn H, editores. Handbook of physiology: A critical, comprehensive presentation of physiological knowledge and concepts. Section 3. Respiration. Washington: American Physiological Society; 1964. p. 463-76.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874679&pid=S0210-5691200900060000300014&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">15. Younes M, Puddy A, Roberts D, Light B, Quesada A, Taylor K, et al. Proportional assist ventilation. Results of an initial clinical trial. Am Rev Respir Dis. 1992;145:121-9.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874680&pid=S0210-5691200900060000300015&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">16. D'Angelo E, Calderini G, Torri G, Robatto FM, Bono D, Milic-Emili J. Respiratory mechanics in anesthetized paralyzed humans: effects of flor, volume, and time. J Appl Physiol. 1989;67: 2556-64.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874681&pid=S0210-5691200900060000300016&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">17. Appendini L, Purro A, Gudjonsdottir M, Baderna P, Tatessio A, Zanoboni S, et al. Physiologic response of ventilator-dependent patients with chronic obstructive pulmonary disease to proportional assist ventilation and continuous positive airway pressure. Am J Respir Crit Care Med. 1999;159:1510-7.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874682&pid=S0210-5691200900060000300017&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">18. Gallagher CG, Younes M. Effect of pressure assist on ventilation and respiratory mechanics in heavy exercise. J Appl Physiol. 1989;66:1824-37.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874683&pid=S0210-5691200900060000300018&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">19. Grasso S, Puntillo F, Mascia L, Ancona G, Fiore T, Bruno F, et al. Compensation for increase in respiratory workload during mechanical ventilation. Pressure-support versus proportional-assist ventilation. Am J Respir Crit Care Med. 2000;161:819-26.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874684&pid=S0210-5691200900060000300019&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">20. Marantz S, Patrick W, Webster K, Roberts D, Oppenheimer L, Younes M. Response of ventilator-dependent patients to different levels of proportional assist. J Appl Physiol. 1996;80:397-403.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874685&pid=S0210-5691200900060000300020&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">21. Bonetto C, Caló MN, Delgado MO, Mancebo J. Modes of pressure delivery and patient-ventilator interaction. Respir Care Clin. 2005;11:247-63.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874686&pid=S0210-5691200900060000300021&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">22. Pérez M, Mancebo J. Monitorización de la mecánica ventilatoria. Med Intensiva. 2006;9:440-8.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874687&pid=S0210-5691200900060000300022&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">23. Wright PE, Marini JJ, Bernard GR. In vitro versus in vivo comparison of endotracheal tube airflow resistance. Am Rev Respir Dis. 1989;140:10-6.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874688&pid=S0210-5691200900060000300023&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">24. Giannouli E, Webster K, Roberts D, Younes M. Response of ventilator dependent patients to different levels of pressure support and proportional assist. Am J Respir Crit Care Med. 1999;159:1716-25.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874689&pid=S0210-5691200900060000300024&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">25. Ranieri VM, Giuliani R, Mascia L, Grasso S, Petruzzelli V, Puntillo N, et al. Patient-ventilator interaction during acute hyper-capnia: pressure-support vs. proportional-assist ventilation. J Appl Physiol. 1996:81:426-36.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874690&pid=S0210-5691200900060000300025&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">26. Mitrouska J, Xirouchaki N, Patakas D, Siafakas N, Georgopoulos D. Effects of chemical feedback on respiratory motor and ventilatory output during different modes of assisted mechanical ventilation. Eur Respir J. 1999;13:873-82.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874691&pid=S0210-5691200900060000300026&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">27. Kondili E, Xirouchaki N, Vaporidi K, Kimathianaki M, Georgopoulos D. Short-term cardiorespiratory effects of proportional assist and pressure-support ventilation in patients with acute lung injury/acute respiratory distress syndrome. Anesthesiology. 2006;105:703-8.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874692&pid=S0210-5691200900060000300027&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">28. Chiumello D, Pelosi P, Groci M, et al. Different modes of assisted ventilation in patients with acute respiratory failure. Eur Respir J. 2002;20:925-33.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874693&pid=S0210-5691200900060000300028&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">29. Brochard L, Harf A, Lorino H, Lemaire F. Inspiratory pressure support prevents diaphragmatic fatigue during weaning from mechanical ventilation. Am Rev Respir Dis. 1989;139:513-21.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874694&pid=S0210-5691200900060000300029&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">30. Clark FJ, Euler C. On the regulation of depth and rate of breathing. J Physiol. 1972;222:267-95.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874695&pid=S0210-5691200900060000300030&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">31. Corne S, Gillespie D, Roberts D, Younes M. Effect of inspiratory flow rate on respiratory rate in intubated ventilated patients. Am J Respir Crit Care Med. 1997;156:304-8.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874696&pid=S0210-5691200900060000300031&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">32. Wilson PA, Skatrud JB, Dempsey JA. Effects of slow wave sleep on ventilatory compensation to inspiratory elastic loading. Respir Physiol. 1984;55:103-20.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874697&pid=S0210-5691200900060000300032&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">33. Navalesi P, Hernandez P, Wongsa A, Laporta D, Goldberg P, Gottfried SB. Proportional assist ventilation in acute respiratory failure: effects on breathing pattern and inspiratory effort. Am J Respir Crit Care Med. 1996;154:1330-8.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874698&pid=S0210-5691200900060000300033&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">34. Iotti GA, Braschi A, Brunner JX, Palo A, Olivei MC. Non invasive evaluation of instantaneous total mechanical activity of the respiratory muscles during pressure support ventilation. Chest. 1995;108:208-15.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874699&pid=S0210-5691200900060000300034&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">35. Yamada Y, Shigeta M, Suwa K, Hanaoka K. Respiratory muscle pressure analysis in pressure-support ventilation. J Appl Physiol. 1994;77:2237-43.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874700&pid=S0210-5691200900060000300035&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">36. Rossi A, Gottfried SB, Zocchi L, Higgs BD, Lennox S, Calverley PM, et al. Measurement of static of the total respiratory system in patients with acute respiratory failure during mechanical ventilation. Effect of intrinsic positive end-expiratory pressure. Am Rev Respir Dis. 1985;131:672-7.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874701&pid=S0210-5691200900060000300036&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">37. Tobin MJ, Van de Graaff WB. Monitoring of lung mechanics and work of breathing. En: Tobin MJ, editor. Principles and practice of mechanical ventilation. New York: MacGraw-Hill; 1994. p. 967-1003.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874702&pid=S0210-5691200900060000300037&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">38. Du HL, Ohtsuji M, Shigeta M, Chao DC, Sasaki K, Usada Y, et al. Expiratory asynchrony in proportional assist ventilation. Am J Respir Crit Care Med. 2002;165:972-7.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874703&pid=S0210-5691200900060000300038&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><!-- ref --><p><font face="Verdana" size="2">39. Passam F, Hoing S, Prinianakis G, Siafakas N, Milic-Emili J, Georgopoulos D. Effect of different levels of pressure support and proportional assist ventilation on breathing pattern, work of breathing and gas exchange in mechanically ventilated hypercapnic COPD patients with acute repiratory failure. Respiration. 2003;70:355-61.</font>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=2874704&pid=S0210-5691200900060000300039&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --><p>&nbsp;</p>     <p>&nbsp;</p>     <p><a name="bajo"></a><a href="#top"><img border="0" src="/img/revistas/medinte/v33n6/seta.gif" width="15" height="17"></a><font face="Verdana" size="2"><b>Dirección para correspondencia:</b>    <br>Correo electrónico: <a href="mailto:frferron@terra.es">frferron@terra.es</a>    <br>(F. Ruiz Ferrón).</font></p>     <p><font face="Verdana" size="2">Recibido el 7 de abril de 2008;    ]]></body>
<body><![CDATA[<br>Aceptado el 22 de diciembre de 2008</font></p>      ]]></body><back>
<ref-list>
<ref id="B1">
<label>1</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Esteban]]></surname>
<given-names><![CDATA[A]]></given-names>
</name>
<name>
<surname><![CDATA[Alía]]></surname>
<given-names><![CDATA[I]]></given-names>
</name>
<name>
<surname><![CDATA[Ibañez]]></surname>
<given-names><![CDATA[J]]></given-names>
</name>
<name>
<surname><![CDATA[Benito]]></surname>
<given-names><![CDATA[S]]></given-names>
</name>
<name>
<surname><![CDATA[Tobin]]></surname>
<given-names><![CDATA[MJ]]></given-names>
</name>
</person-group>
<collab>Spanish lung failure collaborative group</collab>
<article-title xml:lang="en"><![CDATA[Modes of mechanical ventilation and weaning: A national survey of Spanish hospitals]]></article-title>
<source><![CDATA[Chest]]></source>
<year>1994</year>
<volume>106</volume>
<page-range>1188-93</page-range></nlm-citation>
</ref>
<ref id="B2">
<label>2</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Racca]]></surname>
<given-names><![CDATA[F]]></given-names>
</name>
<name>
<surname><![CDATA[Squadrone]]></surname>
<given-names><![CDATA[V]]></given-names>
</name>
<name>
<surname><![CDATA[Ranieri]]></surname>
<given-names><![CDATA[MV]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Patient-vetilator interaction during the triggering phase]]></article-title>
<source><![CDATA[Respir Care Clin.]]></source>
<year>2005</year>
<volume>11</volume>
<page-range>225-45</page-range></nlm-citation>
</ref>
<ref id="B3">
<label>3</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Thille]]></surname>
<given-names><![CDATA[AW]]></given-names>
</name>
<name>
<surname><![CDATA[Rodríguez]]></surname>
<given-names><![CDATA[P]]></given-names>
</name>
<name>
<surname><![CDATA[Cabello]]></surname>
<given-names><![CDATA[B]]></given-names>
</name>
<name>
<surname><![CDATA[Lellouche]]></surname>
<given-names><![CDATA[F]]></given-names>
</name>
<name>
<surname><![CDATA[Brochard]]></surname>
<given-names><![CDATA[L]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Patient-ventilator asynchrony during assisted mechanical ventilation]]></article-title>
<source><![CDATA[Intensive Care Med.]]></source>
<year>2006</year>
<volume>32</volume>
<page-range>1515-22</page-range></nlm-citation>
</ref>
<ref id="B4">
<label>4</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Gay]]></surname>
<given-names><![CDATA[PC]]></given-names>
</name>
<name>
<surname><![CDATA[Hess]]></surname>
<given-names><![CDATA[DR]]></given-names>
</name>
<name>
<surname><![CDATA[Hill]]></surname>
<given-names><![CDATA[NS]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Noninvasive proportional assist ventilation for acute respiratory insufficiency: Comparison with pressure support ventilation]]></article-title>
<source><![CDATA[Am J Respir Crit Care Med.]]></source>
<year>2001</year>
<volume>164</volume>
<page-range>1606-11</page-range></nlm-citation>
</ref>
<ref id="B5">
<label>5</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Meza]]></surname>
<given-names><![CDATA[S]]></given-names>
</name>
<name>
<surname><![CDATA[Younes]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Ventilatory stability during sleep studied with proportional assist ventilation (PAV)]]></article-title>
<source><![CDATA[Sleep.]]></source>
<year>1996</year>
<volume>19</volume>
<page-range>S166</page-range></nlm-citation>
</ref>
<ref id="B6">
<label>6</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Bigatello]]></surname>
<given-names><![CDATA[LM]]></given-names>
</name>
<name>
<surname><![CDATA[Nishimura]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
<name>
<surname><![CDATA[Imanaka]]></surname>
<given-names><![CDATA[H]]></given-names>
</name>
<name>
<surname><![CDATA[Hess]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
<name>
<surname><![CDATA[Kimball]]></surname>
<given-names><![CDATA[WR]]></given-names>
</name>
<name>
<surname><![CDATA[Kacmarek]]></surname>
<given-names><![CDATA[RM]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Unloading of the work of breathing by proportional assist ventilation in a lung model]]></article-title>
<source><![CDATA[Crit Care Med.]]></source>
<year>1997</year>
<volume>25</volume>
<page-range>267-72</page-range></nlm-citation>
</ref>
<ref id="B7">
<label>7</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Delaere]]></surname>
<given-names><![CDATA[S]]></given-names>
</name>
<name>
<surname><![CDATA[Roeseler]]></surname>
<given-names><![CDATA[J]]></given-names>
</name>
<name>
<surname><![CDATA[D'hoore]]></surname>
<given-names><![CDATA[W]]></given-names>
</name>
<name>
<surname><![CDATA[Matte]]></surname>
<given-names><![CDATA[P]]></given-names>
</name>
<name>
<surname><![CDATA[Reynaert]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
<name>
<surname><![CDATA[Jolliet]]></surname>
<given-names><![CDATA[P]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Respiratory muscle workload in intubated, spontaneous breathing without COPD: pressure support vs proportional assist ventilation]]></article-title>
<source><![CDATA[Intensive Care Med.]]></source>
<year>2003</year>
<volume>29</volume>
<page-range>949-54</page-range></nlm-citation>
</ref>
<ref id="B8">
<label>8</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Younes]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Proportional assist ventilation, a new approach to ventilatory support]]></article-title>
<source><![CDATA[Theory. Am Rev Respir Dis.]]></source>
<year>1992</year>
<volume>145</volume>
<page-range>114-20</page-range></nlm-citation>
</ref>
<ref id="B9">
<label>9</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Banner]]></surname>
<given-names><![CDATA[MJ]]></given-names>
</name>
<name>
<surname><![CDATA[Kirby]]></surname>
<given-names><![CDATA[RR]]></given-names>
</name>
<name>
<surname><![CDATA[Kirton]]></surname>
<given-names><![CDATA[OC]]></given-names>
</name>
<name>
<surname><![CDATA[Dehaven]]></surname>
<given-names><![CDATA[B]]></given-names>
</name>
<name>
<surname><![CDATA[Blanch]]></surname>
<given-names><![CDATA[PB]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Breathing frequency and pattern are poor predictors of work of breathing in patients receiving pressure support ventilation]]></article-title>
<source><![CDATA[Chest.]]></source>
<year>1995</year>
<volume>108</volume>
<page-range>1338-44</page-range></nlm-citation>
</ref>
<ref id="B10">
<label>10</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Younes]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
<name>
<surname><![CDATA[Kun]]></surname>
<given-names><![CDATA[J]]></given-names>
</name>
<name>
<surname><![CDATA[Masiowski]]></surname>
<given-names><![CDATA[B]]></given-names>
</name>
<name>
<surname><![CDATA[Webster]]></surname>
<given-names><![CDATA[K]]></given-names>
</name>
<name>
<surname><![CDATA[Roberts]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[A method for non-invasive determination of inspiratory resistance during proportional assist ventilation]]></article-title>
<source><![CDATA[Am J Respir Crit Care Med.]]></source>
<year>2001</year>
<volume>163</volume>
<page-range>829-39</page-range></nlm-citation>
</ref>
<ref id="B11">
<label>11</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Younes]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
<name>
<surname><![CDATA[Webster]]></surname>
<given-names><![CDATA[K]]></given-names>
</name>
<name>
<surname><![CDATA[Kun]]></surname>
<given-names><![CDATA[J]]></given-names>
</name>
<name>
<surname><![CDATA[Roberts]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
<name>
<surname><![CDATA[Masiowski]]></surname>
<given-names><![CDATA[T]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[A method for measuring passive elastance during proportional assist ventilation]]></article-title>
<source><![CDATA[Am J Respir Crit Care Med.]]></source>
<year>2001</year>
<volume>164</volume>
<page-range>50-60</page-range></nlm-citation>
</ref>
<ref id="B12">
<label>12</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Younes]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
<name>
<surname><![CDATA[Kun]]></surname>
<given-names><![CDATA[J]]></given-names>
</name>
<name>
<surname><![CDATA[Webster]]></surname>
<given-names><![CDATA[K]]></given-names>
</name>
<name>
<surname><![CDATA[Roberts]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Response of ventilator-dependent patients to delayed opening of exhalation valve]]></article-title>
<source><![CDATA[Am J Respir Crit Care Med.]]></source>
<year>2002</year>
<volume>166</volume>
<page-range>21-30</page-range></nlm-citation>
</ref>
<ref id="B13">
<label>13</label><nlm-citation citation-type="book">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Younes]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Proportional assist ventilation]]></article-title>
<person-group person-group-type="editor">
<name>
<surname><![CDATA[Tobin]]></surname>
<given-names><![CDATA[MJ]]></given-names>
</name>
</person-group>
<source><![CDATA[Principles and practice of mechanical ventilation]]></source>
<year>1994</year>
<page-range>349-69</page-range><publisher-loc><![CDATA[New York ]]></publisher-loc>
<publisher-name><![CDATA[MacGraw-Hill]]></publisher-name>
</nlm-citation>
</ref>
<ref id="B14">
<label>14</label><nlm-citation citation-type="book">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Otis]]></surname>
<given-names><![CDATA[AB]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[The work of breathing]]></article-title>
<person-group person-group-type="editor">
<name>
<surname><![CDATA[Fenn]]></surname>
<given-names><![CDATA[WO]]></given-names>
</name>
<name>
<surname><![CDATA[Rahn]]></surname>
<given-names><![CDATA[H]]></given-names>
</name>
</person-group>
<source><![CDATA[Handbook of physiology: A critical, comprehensive presentation of physiological knowledge and concepts. Section 3. Respiration]]></source>
<year>1964</year>
<page-range>463-76</page-range><publisher-loc><![CDATA[Washington ]]></publisher-loc>
<publisher-name><![CDATA[American Physiological Society]]></publisher-name>
</nlm-citation>
</ref>
<ref id="B15">
<label>15</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Younes]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
<name>
<surname><![CDATA[Puddy]]></surname>
<given-names><![CDATA[A]]></given-names>
</name>
<name>
<surname><![CDATA[Roberts]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
<name>
<surname><![CDATA[Light]]></surname>
<given-names><![CDATA[B]]></given-names>
</name>
<name>
<surname><![CDATA[Quesada]]></surname>
<given-names><![CDATA[A]]></given-names>
</name>
<name>
<surname><![CDATA[Taylor]]></surname>
<given-names><![CDATA[K]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Proportional assist ventilation: Results of an initial clinical trial]]></article-title>
<source><![CDATA[Am Rev Respir Dis.]]></source>
<year>1992</year>
<volume>145</volume>
<page-range>121-9</page-range></nlm-citation>
</ref>
<ref id="B16">
<label>16</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[D'Angelo]]></surname>
<given-names><![CDATA[E]]></given-names>
</name>
<name>
<surname><![CDATA[Calderini]]></surname>
<given-names><![CDATA[G]]></given-names>
</name>
<name>
<surname><![CDATA[Torri]]></surname>
<given-names><![CDATA[G]]></given-names>
</name>
<name>
<surname><![CDATA[Robatto]]></surname>
<given-names><![CDATA[FM]]></given-names>
</name>
<name>
<surname><![CDATA[Bono]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
<name>
<surname><![CDATA[Milic-Emili]]></surname>
<given-names><![CDATA[J]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Respiratory mechanics in anesthetized paralyzed humans: effects of flor, volume, and time]]></article-title>
<source><![CDATA[J Appl Physiol.]]></source>
<year>1989</year>
<volume>67</volume>
<page-range>2556-64</page-range></nlm-citation>
</ref>
<ref id="B17">
<label>17</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Appendini]]></surname>
<given-names><![CDATA[L]]></given-names>
</name>
<name>
<surname><![CDATA[Purro]]></surname>
<given-names><![CDATA[A]]></given-names>
</name>
<name>
<surname><![CDATA[Gudjonsdottir]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
<name>
<surname><![CDATA[Baderna]]></surname>
<given-names><![CDATA[P]]></given-names>
</name>
<name>
<surname><![CDATA[Tatessio]]></surname>
<given-names><![CDATA[A]]></given-names>
</name>
<name>
<surname><![CDATA[Zanoboni]]></surname>
<given-names><![CDATA[S]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Physiologic response of ventilator-dependent patients with chronic obstructive pulmonary disease to proportional assist ventilation and continuous positive airway pressure]]></article-title>
<source><![CDATA[Am J Respir Crit Care Med.]]></source>
<year>1999</year>
<volume>159</volume>
<page-range>1510-7</page-range></nlm-citation>
</ref>
<ref id="B18">
<label>18</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Gallagher]]></surname>
<given-names><![CDATA[CG]]></given-names>
</name>
<name>
<surname><![CDATA[Younes]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Effect of pressure assist on ventilation and respiratory mechanics in heavy exercise]]></article-title>
<source><![CDATA[J Appl Physiol.]]></source>
<year>1989</year>
<volume>66</volume>
<page-range>1824-37</page-range></nlm-citation>
</ref>
<ref id="B19">
<label>19</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Grasso]]></surname>
<given-names><![CDATA[S]]></given-names>
</name>
<name>
<surname><![CDATA[Puntillo]]></surname>
<given-names><![CDATA[F]]></given-names>
</name>
<name>
<surname><![CDATA[Mascia]]></surname>
<given-names><![CDATA[L]]></given-names>
</name>
<name>
<surname><![CDATA[Ancona]]></surname>
<given-names><![CDATA[G]]></given-names>
</name>
<name>
<surname><![CDATA[Fiore]]></surname>
<given-names><![CDATA[T]]></given-names>
</name>
<name>
<surname><![CDATA[Bruno]]></surname>
<given-names><![CDATA[F]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Compensation for increase in respiratory workload during mechanical ventilation: Pressure-support versus proportional-assist ventilation]]></article-title>
<source><![CDATA[Am J Respir Crit Care Med.]]></source>
<year>2000</year>
<volume>161</volume>
<page-range>819-26</page-range></nlm-citation>
</ref>
<ref id="B20">
<label>20</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Marantz]]></surname>
<given-names><![CDATA[S]]></given-names>
</name>
<name>
<surname><![CDATA[Patrick]]></surname>
<given-names><![CDATA[W]]></given-names>
</name>
<name>
<surname><![CDATA[Webster]]></surname>
<given-names><![CDATA[K]]></given-names>
</name>
<name>
<surname><![CDATA[Roberts]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
<name>
<surname><![CDATA[Oppenheimer]]></surname>
<given-names><![CDATA[L]]></given-names>
</name>
<name>
<surname><![CDATA[Younes]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Response of ventilator-dependent patients to different levels of proportional assist]]></article-title>
<source><![CDATA[J Appl Physiol.]]></source>
<year>1996</year>
<volume>80</volume>
<page-range>397-403</page-range></nlm-citation>
</ref>
<ref id="B21">
<label>21</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Bonetto]]></surname>
<given-names><![CDATA[C]]></given-names>
</name>
<name>
<surname><![CDATA[Caló]]></surname>
<given-names><![CDATA[MN]]></given-names>
</name>
<name>
<surname><![CDATA[Delgado]]></surname>
<given-names><![CDATA[MO]]></given-names>
</name>
<name>
<surname><![CDATA[Mancebo]]></surname>
<given-names><![CDATA[J]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Modes of pressure delivery and patient-ventilator interaction]]></article-title>
<source><![CDATA[Respir Care Clin.]]></source>
<year>2005</year>
<volume>11</volume>
<page-range>247-63</page-range></nlm-citation>
</ref>
<ref id="B22">
<label>22</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Pérez]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
<name>
<surname><![CDATA[Mancebo]]></surname>
<given-names><![CDATA[J]]></given-names>
</name>
</person-group>
<article-title xml:lang="es"><![CDATA[Monitorización de la mecánica ventilatoria]]></article-title>
<source><![CDATA[Med Intensiva.]]></source>
<year>2006</year>
<volume>9</volume>
<page-range>440-8</page-range></nlm-citation>
</ref>
<ref id="B23">
<label>23</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Wright]]></surname>
<given-names><![CDATA[PE]]></given-names>
</name>
<name>
<surname><![CDATA[Marini]]></surname>
<given-names><![CDATA[JJ]]></given-names>
</name>
<name>
<surname><![CDATA[Bernard]]></surname>
<given-names><![CDATA[GR]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[In vitro versus in vivo comparison of endotracheal tube airflow resistance]]></article-title>
<source><![CDATA[Am Rev Respir Dis.]]></source>
<year>1989</year>
<volume>140</volume>
<page-range>10-6</page-range></nlm-citation>
</ref>
<ref id="B24">
<label>24</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Giannouli]]></surname>
<given-names><![CDATA[E]]></given-names>
</name>
<name>
<surname><![CDATA[Webster]]></surname>
<given-names><![CDATA[K]]></given-names>
</name>
<name>
<surname><![CDATA[Roberts]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
<name>
<surname><![CDATA[Younes]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Response of ventilator dependent patients to different levels of pressure support and proportional assist]]></article-title>
<source><![CDATA[Am J Respir Crit Care Med.]]></source>
<year>1999</year>
<volume>159</volume>
<page-range>1716-25</page-range></nlm-citation>
</ref>
<ref id="B25">
<label>25</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Ranieri]]></surname>
<given-names><![CDATA[VM]]></given-names>
</name>
<name>
<surname><![CDATA[Giuliani]]></surname>
<given-names><![CDATA[R]]></given-names>
</name>
<name>
<surname><![CDATA[Mascia]]></surname>
<given-names><![CDATA[L]]></given-names>
</name>
<name>
<surname><![CDATA[Grasso]]></surname>
<given-names><![CDATA[S]]></given-names>
</name>
<name>
<surname><![CDATA[Petruzzelli]]></surname>
<given-names><![CDATA[V]]></given-names>
</name>
<name>
<surname><![CDATA[Puntillo]]></surname>
<given-names><![CDATA[N]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Patient-ventilator interaction during acute hyper-capnia: pressure-support vs. proportional-assist ventilation]]></article-title>
<source><![CDATA[J Appl Physiol.]]></source>
<year>1996</year>
<volume>81</volume>
<page-range>426-36</page-range></nlm-citation>
</ref>
<ref id="B26">
<label>26</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Mitrouska]]></surname>
<given-names><![CDATA[J]]></given-names>
</name>
<name>
<surname><![CDATA[Xirouchaki]]></surname>
<given-names><![CDATA[N]]></given-names>
</name>
<name>
<surname><![CDATA[Patakas]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
<name>
<surname><![CDATA[Siafakas]]></surname>
<given-names><![CDATA[N]]></given-names>
</name>
<name>
<surname><![CDATA[Georgopoulos]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Effects of chemical feedback on respiratory motor and ventilatory output during different modes of assisted mechanical ventilation]]></article-title>
<source><![CDATA[Eur Respir J.]]></source>
<year>1999</year>
<volume>13</volume>
<page-range>873-82</page-range></nlm-citation>
</ref>
<ref id="B27">
<label>27</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Kondili]]></surname>
<given-names><![CDATA[E]]></given-names>
</name>
<name>
<surname><![CDATA[Xirouchaki]]></surname>
<given-names><![CDATA[N]]></given-names>
</name>
<name>
<surname><![CDATA[Vaporidi]]></surname>
<given-names><![CDATA[K]]></given-names>
</name>
<name>
<surname><![CDATA[Kimathianaki]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
<name>
<surname><![CDATA[Georgopoulos]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Short-term cardiorespiratory effects of proportional assist and pressure-support ventilation in patients with acute lung injury/acute respiratory distress syndrome]]></article-title>
<source><![CDATA[Anesthesiology.]]></source>
<year>2006</year>
<volume>105</volume>
<page-range>703-8</page-range></nlm-citation>
</ref>
<ref id="B28">
<label>28</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Chiumello]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
<name>
<surname><![CDATA[Pelosi]]></surname>
<given-names><![CDATA[P]]></given-names>
</name>
<name>
<surname><![CDATA[Groci]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Different modes of assisted ventilation in patients with acute respiratory failure]]></article-title>
<source><![CDATA[Eur Respir J.]]></source>
<year>2002</year>
<volume>20</volume>
<page-range>925-33</page-range></nlm-citation>
</ref>
<ref id="B29">
<label>29</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Brochard]]></surname>
<given-names><![CDATA[L]]></given-names>
</name>
<name>
<surname><![CDATA[Harf]]></surname>
<given-names><![CDATA[A]]></given-names>
</name>
<name>
<surname><![CDATA[Lorino]]></surname>
<given-names><![CDATA[H]]></given-names>
</name>
<name>
<surname><![CDATA[Lemaire]]></surname>
<given-names><![CDATA[F]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Inspiratory pressure support prevents diaphragmatic fatigue during weaning from mechanical ventilation]]></article-title>
<source><![CDATA[Am Rev Respir Dis.]]></source>
<year>1989</year>
<volume>139</volume>
<page-range>513-21</page-range></nlm-citation>
</ref>
<ref id="B30">
<label>30</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Clark]]></surname>
<given-names><![CDATA[FJ]]></given-names>
</name>
<name>
<surname><![CDATA[Euler]]></surname>
<given-names><![CDATA[C]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[On the regulation of depth and rate of breathing]]></article-title>
<source><![CDATA[J Physiol.]]></source>
<year>1972</year>
<volume>222</volume>
<page-range>267-95</page-range></nlm-citation>
</ref>
<ref id="B31">
<label>31</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Corne]]></surname>
<given-names><![CDATA[S]]></given-names>
</name>
<name>
<surname><![CDATA[Gillespie]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
<name>
<surname><![CDATA[Roberts]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
<name>
<surname><![CDATA[Younes]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Effect of inspiratory flow rate on respiratory rate in intubated ventilated patients]]></article-title>
<source><![CDATA[Am J Respir Crit Care Med.]]></source>
<year>1997</year>
<volume>156</volume>
<page-range>304-8</page-range></nlm-citation>
</ref>
<ref id="B32">
<label>32</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Wilson]]></surname>
<given-names><![CDATA[PA]]></given-names>
</name>
<name>
<surname><![CDATA[Skatrud]]></surname>
<given-names><![CDATA[JB]]></given-names>
</name>
<name>
<surname><![CDATA[Dempsey]]></surname>
<given-names><![CDATA[JA]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Effects of slow wave sleep on ventilatory compensation to inspiratory elastic loading]]></article-title>
<source><![CDATA[Respir Physiol.]]></source>
<year>1984</year>
<volume>55</volume>
<page-range>103-20</page-range></nlm-citation>
</ref>
<ref id="B33">
<label>33</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Navalesi]]></surname>
<given-names><![CDATA[P]]></given-names>
</name>
<name>
<surname><![CDATA[Hernandez]]></surname>
<given-names><![CDATA[P]]></given-names>
</name>
<name>
<surname><![CDATA[Wongsa]]></surname>
<given-names><![CDATA[A]]></given-names>
</name>
<name>
<surname><![CDATA[Laporta]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
<name>
<surname><![CDATA[Goldberg]]></surname>
<given-names><![CDATA[P]]></given-names>
</name>
<name>
<surname><![CDATA[Gottfried]]></surname>
<given-names><![CDATA[SB]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Proportional assist ventilation in acute respiratory failure: effects on breathing pattern and inspiratory effort]]></article-title>
<source><![CDATA[Am J Respir Crit Care Med.]]></source>
<year>1996</year>
<volume>154</volume>
<page-range>1330-8</page-range></nlm-citation>
</ref>
<ref id="B34">
<label>34</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Iotti]]></surname>
<given-names><![CDATA[GA]]></given-names>
</name>
<name>
<surname><![CDATA[Braschi]]></surname>
<given-names><![CDATA[A]]></given-names>
</name>
<name>
<surname><![CDATA[Brunner]]></surname>
<given-names><![CDATA[JX]]></given-names>
</name>
<name>
<surname><![CDATA[Palo]]></surname>
<given-names><![CDATA[A]]></given-names>
</name>
<name>
<surname><![CDATA[Olivei]]></surname>
<given-names><![CDATA[MC]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Non invasive evaluation of instantaneous total mechanical activity of the respiratory muscles during pressure support ventilation]]></article-title>
<source><![CDATA[Chest.]]></source>
<year>1995</year>
<volume>108</volume>
<page-range>208-15</page-range></nlm-citation>
</ref>
<ref id="B35">
<label>35</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Yamada]]></surname>
<given-names><![CDATA[Y]]></given-names>
</name>
<name>
<surname><![CDATA[Shigeta]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
<name>
<surname><![CDATA[Suwa]]></surname>
<given-names><![CDATA[K]]></given-names>
</name>
<name>
<surname><![CDATA[Hanaoka]]></surname>
<given-names><![CDATA[K]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Respiratory muscle pressure analysis in pressure-support ventilation]]></article-title>
<source><![CDATA[J Appl Physiol.]]></source>
<year>1994</year>
<volume>77</volume>
<page-range>2237-43</page-range></nlm-citation>
</ref>
<ref id="B36">
<label>36</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Rossi]]></surname>
<given-names><![CDATA[A]]></given-names>
</name>
<name>
<surname><![CDATA[Gottfried]]></surname>
<given-names><![CDATA[SB]]></given-names>
</name>
<name>
<surname><![CDATA[Zocchi]]></surname>
<given-names><![CDATA[L]]></given-names>
</name>
<name>
<surname><![CDATA[Higgs]]></surname>
<given-names><![CDATA[BD]]></given-names>
</name>
<name>
<surname><![CDATA[Lennox]]></surname>
<given-names><![CDATA[S]]></given-names>
</name>
<name>
<surname><![CDATA[Calverley]]></surname>
<given-names><![CDATA[PM]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Measurement of static of the total respiratory system in patients with acute respiratory failure during mechanical ventilation: Effect of intrinsic positive end-expiratory pressure]]></article-title>
<source><![CDATA[Am Rev Respir Dis.]]></source>
<year>1985</year>
<volume>131</volume>
<page-range>672-7</page-range></nlm-citation>
</ref>
<ref id="B37">
<label>37</label><nlm-citation citation-type="book">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Tobin]]></surname>
<given-names><![CDATA[MJ]]></given-names>
</name>
<name>
<surname><![CDATA[Van de Graaff]]></surname>
<given-names><![CDATA[WB]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Monitoring of lung mechanics and work of breathing]]></article-title>
<person-group person-group-type="editor">
<name>
<surname><![CDATA[Tobin]]></surname>
<given-names><![CDATA[MJ]]></given-names>
</name>
</person-group>
<source><![CDATA[Principles and practice of mechanical ventilation]]></source>
<year>1994</year>
<page-range>967-1003</page-range><publisher-loc><![CDATA[New York ]]></publisher-loc>
<publisher-name><![CDATA[MacGraw-Hill]]></publisher-name>
</nlm-citation>
</ref>
<ref id="B38">
<label>38</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Du]]></surname>
<given-names><![CDATA[HL]]></given-names>
</name>
<name>
<surname><![CDATA[Ohtsuji]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
<name>
<surname><![CDATA[Shigeta]]></surname>
<given-names><![CDATA[M]]></given-names>
</name>
<name>
<surname><![CDATA[Chao]]></surname>
<given-names><![CDATA[DC]]></given-names>
</name>
<name>
<surname><![CDATA[Sasaki]]></surname>
<given-names><![CDATA[K]]></given-names>
</name>
<name>
<surname><![CDATA[Usada]]></surname>
<given-names><![CDATA[Y]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Expiratory asynchrony in proportional assist ventilation]]></article-title>
<source><![CDATA[Am J Respir Crit Care Med.]]></source>
<year>2002</year>
<volume>165</volume>
<page-range>972-7</page-range></nlm-citation>
</ref>
<ref id="B39">
<label>39</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Passam]]></surname>
<given-names><![CDATA[F]]></given-names>
</name>
<name>
<surname><![CDATA[Hoing]]></surname>
<given-names><![CDATA[S]]></given-names>
</name>
<name>
<surname><![CDATA[Prinianakis]]></surname>
<given-names><![CDATA[G]]></given-names>
</name>
<name>
<surname><![CDATA[Siafakas]]></surname>
<given-names><![CDATA[N]]></given-names>
</name>
<name>
<surname><![CDATA[Milic-Emili]]></surname>
<given-names><![CDATA[J]]></given-names>
</name>
<name>
<surname><![CDATA[Georgopoulos]]></surname>
<given-names><![CDATA[D]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Effect of different levels of pressure support and proportional assist ventilation on breathing pattern, work of breathing and gas exchange in mechanically ventilated hypercapnic COPD patients with acute repiratory failure]]></article-title>
<source><![CDATA[Respiration.]]></source>
<year>2003</year>
<volume>70</volume>
<page-range>355-61</page-range></nlm-citation>
</ref>
</ref-list>
</back>
</article>